ZAN Zhi-sheng,QIN Wei-chang,QI Heng-tao.Comparison of low-density signal detection between CR and DR systems[J].Chinese Journal of Radiological Medicine and Protection,2006,26(6):617-618 |
Comparison of low-density signal detection between CR and DR systems |
Received:January 10, 2006 |
DOI: |
KeyWords:Computed radiography Digital radiography Detection capability |
FundProject: |
|
Hits: 2934 |
Download times: 1960 |
Abstract:: |
Objective To compare the low-density signal detection between CR and DR systems. Methods Burger's phantom (Kyoto Scientific Corporation), 20 mm aluminium plate (RPM X-QA, Beijing Institute of Radiation Medicine), 0.5 mm copper plate, RTI Solidose 400 digital dosimeter, PMX-IR kV-time meter, and CPI high-frequency generator were applied to compare the low density signal detection between CR and DR systems for the same X-ray machine, absorber and exposed dose. Results The low-density signal detection capability is 6.5% less than that of the DR system. Conclusions CR system is inferior to DR system (amorphous selenium flat plate panels) in low-density signal detection. |