ZHAO Song,ZHANG Jun-yi,ZHANG Wei-min.Low-dose investigation of digital radiography system image[J].Chinese Journal of Radiological Medicine and Protection,2006,26(4):396-398
Low-dose investigation of digital radiography system image
Received:January 08, 2006  
DOI:
KeyWords:Digital radiography  Contrast-detail phantom  Image quality factor(IQF)  Image quality
FundProject:浙江省卫生科技基金资助项目(2004A039)
Author NameAffiliationE-mail
ZHAO Song Department of Radiology, Seconded Affiliated Hospital, Zhejiang University, Hangzhou 310009, China zwem@mail.hz.zj.cn 
ZHANG Jun-yi 泰山医学院放射系  
ZHANG Wei-min Department of Radiology, Seconded Affiliated Hospital, Zhejiang University, Hangzhou 310009, China  
张德营 Department of Radiology, Seconded Affiliated Hospital, Zhejiang University, Hangzhou 310009, China  
李月卿 泰山医学院放射系  
王昌元 泰山医学院放射系  
侯庆峰 泰山医学院放射系  
Hits: 2576
Download times: 1694
Abstract::
      Objective To evaluate the low dose performance of digital radiography system for the imaging of chest. Methods In this experiment, the radiography of CDRAD 2.0 contrast-detail phantom was done in order to evaluate the relations between image quality and entrance surface dose. Thirty chest radiographys were obtained by CR and DR, the ESD of DR used at 1/3 of CR's. The image definition of medial borders of the scapulae et al was evaluated by six doctors from department of radiology. Results The result indicated no statistically significant different in diagnostic quality between from the investigational system and from CR(P>0.05). Conclusion The superior resolution, low noise characteristics and high DQE of digital radiography can be utilized to decrease the radiation dose exposure to patients, without influencing the diagnostic quality.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9061185  On-line:0

v
Scan QR Code
&et=B70C87E386AAE5F1751538E3DB32B723BAEA60387F8580971783B78D8DA9466FDFBCE1EAB3A086B02E72F196151B2E393F556258B5D57E60428DB9138068E97F&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=37904DC365DD7266&aid=01E7B4173546E258&vid=&iid=E158A972A605785F&sid=7D6CD8918B045FD4&eid=4F0B2F798E08B761&fileno=20060431&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="37904DC365DD7266"; var my_aid="01E7B4173546E258";