WANG Jian_chao,YUE Bao_rong,LIU Lan_tao.Measurement of some influence quantities of the calibration apparatus for mammograhy dosimeters[J].Chinese Journal of Radiological Medicine and Protection,2006,26(3):286-287 |
Measurement of some influence quantities of the calibration apparatus for mammograhy dosimeters |
Received:April 25, 2006 |
DOI: |
KeyWords:Calibration Apparatus Mammography dosimeter Standard radiation qualities Influnce quantitity |
FundProject: |
Author Name | Affiliation | WANG Jian_chao | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | YUE Bao_rong | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | LIU Lan_tao | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | 尉可道 | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China |
|
Hits: 3754 |
Download times: 2101 |
Abstract:: |
Objective To measure some influence quantities of the calibration apparatus for Mammography dosimeters. Methods Under the standard radiation quality conditions,the radiation field homogeneity, the scatter radiation contributions and long term stability of the reference instrument were measured using the reference_class ionization chamber at the SSDL. Results The field homogeneity (Ф60 mm)is in ±0.98%; the scatter radiation contributions at the poit of test were below 0.12%,the long term stability of calibration apparatus was no more than 1.8%(±1 s)over 5 months. Conclusion As indicated above the some influence quantities may meet the requirements for the calibration apparatus at the SSDL. |
HTML View Full Text View/Add Comment Download reader |