CHEN Chao-min,TANG Mu-tao,ZHOU Ling-hong,et al.Study of dose calculation and beam parameters optimization with genetic algorithm in IMRT[J].Chinese Journal of Radiological Medicine and Protection,2006,26(3):248-250
Study of dose calculation and beam parameters optimization with genetic algorithm in IMRT
Received:July 11, 2005  
DOI:
KeyWords:IMRT  Dose calculation  Convolution  Genetic algorithm  Parameters optimization
FundProject:广东省科技计划资助项目(2003C32724);广州市科技攻关资助项目(2003Z3-E0411)
Author NameAffiliationE-mail
CHEN Chao-min School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China gzccm@fimmu.com 
TANG Mu-tao School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China  
ZHOU Ling-hong School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China  
吕庆文 School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China  
王卓宇 School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China  
陈光杰 School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China  
Hits: 3147
Download times: 2602
Abstract::
      Objective To study the construction of dose calculation model and the method of automatic beam parameters selection in IMRT. Methods The three-dimension convolution dose calculation model of photon was constructed with the methods of Fast Fourier Transform. The objective function based on dose constrain was used to evaluate the fitness of individuals. The beam weights were optimized with genetic algorithm. Results After 100 iterative analyses, the treatment planning system produced highly conformal and homogeneous dose distributions. Conclusion the three-dimension convolution dose calculation model of photon gave more accurate results than the conventional models; genetic algorithm is valid and efficient in IMRT beam parameters optimization.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12030300  On-line:0

v
Scan QR Code
&et=1946763CE42DED5D790421BC089FDFEB65630753A25B9624F47FFAB1F5618B3BE9EBB27DFA1B9D241A6F998B20CFE67BFE8C84E3BA197E130F9ED3393DD69AA3DADE12C4B03F40A5B376EBA76BFBD1B5451F1D259FEA83B8C6E2A35071EE70DD&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=37904DC365DD7266&aid=269A7DC1737D89DD&vid=&iid=38B194292C032A66&sid=FBA00558C57D9C11&eid=1D01216AD76577EC&fileno=20060316&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="37904DC365DD7266"; var my_aid="269A7DC1737D89DD";