HU Jia-cheng,YANG Xiao-yuan,YANG Yuan-di,et al.Measurement and verification of wall correction factor for the spherical graphite cavity chamber[J].Chinese Journal of Radiological Medicine and Protection,2005,25(6):561-564
Measurement and verification of wall correction factor for the spherical graphite cavity chamber
Received:June 27, 2005  
DOI:
KeyWords:Cavity chamber  Air kerma  Wall correction factor (Kwall)  Monte Carlo calculations
FundProject:国家自然科学基金资助项目(10275058)
Author NameAffiliation
HU Jia-cheng National Institute of Metrology, Beijing 100013, China 
YANG Xiao-yuan National Institute of Metrology, Beijing 100013, China 
YANG Yuan-di National Institute of Metrology, Beijing 100013, China 
樊铁栓 北京大学物理学院重离子物理研究所重离子物理教育部重点实验室 
黄斐增 北京大学物理学院重离子物理研究所重离子物理教育部重点实验室 
刘鉴常 北京大学物理学院重离子物理研究所重离子物理教育部重点实验室 
孙吉宁 北京大学物理学院重离子物理研究所重离子物理教育部重点实验室 
叶卫国 北京大学物理学院重离子物理研究所重离子物理教育部重点实验室 
Hits: 4015
Download times: 2502
Abstract::
      Objective To discuss the method how to identify wall correction factor (Kwall). Methods The paper introduce how to use extrapolation of the ionization current measured for different wall thickness by the spherical graphite cavity chamber whose volume is 10 cm3 and 30 cm3 for 60Co γ-rays. The wall correction factors were determined by this method and Monte Carlo(MC) calculations. Results The results of the present experiment and calculation indicated that the wall correction factor determined by traditional geometry thickness extrapolation method was 1.0% less than that by MC calculation.The result of applying the equivalent mean thickness to calculate with extrapolation method and that of MC calculation were accordable within 0.45%.The relative dose and wall correction factors for the spherical graphite cavity chamber whose volume is 30 cm3 and 50 cm3 of the standards of air kerma of the NIST for 60Co γ-rays were calculated by MC calculation,and the calculation results were in accordance with the published values of NIST within 0.06%. Conclusion The spherical graphite cavity chamber wall correction factor determined by traditional extrapolation method and MC calculation is accordable with the results of foreign labs (PSDL).
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12035452  On-line:0

v
Scan QR Code
&et=DC44477595533BA79D4991F2B6CA64D02285DDC28546049ADDBFD6D2D66DFB070C4241A3FF78AF5158FAD5BA93055052AD69CA9E337A14038F0418D744812848BEA8ACE9C242BF4DAB6A77583135E036A7635DEFB2C7C51C1BAE2ADB1721C7463A9CDC92B9B5701F&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=2DD7160C83D0ACED&aid=DFBBD65293F96CD4&vid=&iid=B31275AF3241DB2D&sid=B4F9D541F855CF96&eid=2497388423811B81&fileno=20050623&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="2DD7160C83D0ACED"; var my_aid="DFBBD65293F96CD4";