LUO Su-ming,LI Jing-yun,HE Zhi-jian.Development of a well- type chamber for measurement of source activation in breachtherapy[J].Chinese Journal of Radiological Medicine and Protection,2005,25(4):370-373
Development of a well- type chamber for measurement of source activation in breachtherapy
Received:April 24, 2005  
DOI:
KeyWords:Well-type chamber  Air kerma strength of brachytherapy source  Measurement accuracy of source activity
FundProject:北京市自然科学基金资助项目(3042016)4
Author NameAffiliation
LUO Su-ming National Institute for Radiological Protection, China CDC, Beijing 100088, China 
LI Jing-yun 中国原子能科学研究院 
HE Zhi-jian National Institute for Radiological Protection, China CDC, Beijing 100088, China 
朱卫国 National Institute for Radiological Protection, China CDC, Beijing 100088, China 
李开宝 National Institute for Radiological Protection, China CDC, Beijing 100088, China 
刘立明 National Institute for Radiological Protection, China CDC, Beijing 100088, China 
Hits: 2621
Download times: 2138
Abstract::
      Objective To develop a well-type chamber air kerma strength of source for measurement and improve accuracy for source activity. Methods By drawing advanced experience from abroad and combining the condition of our country, we designed a plan, drew the blueprint, selected the material, processed and assembled the well- type chamber, and carried out a performance test. Result The imported and self-made well- type chambers were measured in the same condition for comparision. The results should that the long- term stability of the self-made well- type chamber is 0.4%, and the technique index is 2%; the recombination rate of ionized charges is 0.9995% and the technique index is 1.000; the measurement accuracy of the well-type chamber is 0.02%,and the technique index is 0.5%. For the self-made well-type chamber, the flat response of peak sensitivity is fixed while for imported one the flat response changes around 0.1% within 5 mm of peak sensitivity. Conclusions The strong points of the self-made well-type chamber are quick speed and accuracy in measurement and can measure for 192Ir, 125I and 103Pd sources. Its measurement range is from 3.7 MBq to 7.4 ×105 MBq. The development of the well-type chamber fills the gaps of on -site measurement instruments and becomes a product with independent intellectual property right in our country.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9068226  On-line:0

v
Scan QR Code
&et=A90EAC7419FB9B220C8CA3675965BF0DC0876BE2F7FCFFEC29F75C1F3EF3CC0B4512E92795B59A65409686B2427A5D14D05863860612719BFFBE5B7DDC08ACDB155D82629AE7CE28D222640E33E4E2014FBC02DF78C88FE33411C35E9BDE1F34&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=2DD7160C83D0ACED&aid=6DAD218F301CB86E&vid=&iid=E158A972A605785F&sid=0918129209B14F3E&eid=F4BDB5452F9F5642&fileno=20050430&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="2DD7160C83D0ACED"; var my_aid="6DAD218F301CB86E";