WANG Peng-cheng,ZHANG Fu-li,LI Shi-Jun.Comparison between two different digital radiographic techniques:image quality versus imaging dose[J].Chinese Journal of Radiological Medicine and Protection,2005,25(1):91-93
Comparison between two different digital radiographic techniques:image quality versus imaging dose
Received:April 24, 2004  
DOI:
KeyWords:Digital radiography  Image quality
FundProject:
Author NameAffiliation
WANG Peng-cheng Department of Radiology, Taishan Medical College, Shandong 271000, China 
ZHANG Fu-li Department of Radiology, Taishan Medical College, Shandong 271000, China 
LI Shi-Jun 苏州大学放射医学与公共卫生学院 
Hits: 3902
Download times: 3165
Abstract::
      Objective To assess and compare the difference in image quality and exposure dose between an amorphous selenium flat panel digital radiography(DR) system and computed radiography(CR). Methods A contrast-detail phantom CDRAD2.0 was exposed by DR and CR with different mAs sets.The entrance surface doses were recorded for all images.Images were presented to 4 independent observers on a higher resolution monitor of diagnosis work station.The image quality figure(IQF)was measured for each image.Statistical analysis was performed using ANOVA test to compare the difference in image quality and exposure dose between DR and CR. Results Comparing to CR,DR had a lower IQF,ensuring a high image quality with respect to contrast and detail detectability.When IQFs for DR and CR were equal,DR could reduce the surface dose by 77% of that of CR. Conclusion Image quality of DR has been proved to be far superior to CR,in particular for contrast-detail.The image quality of CR is similar to that of DR only at very high dose levels.The clinical application of DR will reduce the exposure dose to the patient greatly.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12029244  On-line:0

v
Scan QR Code
&et=8C10CEA789A144B857F6EC66CCF7E98C7933D2194DF32BEA610E8B014F41BC94C0D7E00B13245136FBAEEB05E1C3A58E697986683DE37432888D68A58D2297C4A5C6786D81013B7AF03CE13BD00193618B472AC5C8867921291BA379993A218139EDC8BF26C8FD1C&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=2DD7160C83D0ACED&aid=54F9438A1F3747A6&vid=&iid=CA4FD0336C81A37A&sid=C753EB8AC8F551B9&eid=39EEF47180459690&fileno=20050141&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="2DD7160C83D0ACED"; var my_aid="54F9438A1F3747A6";