CHEN Sheng-li,ZHU Dong-liang,CHEN Guo-dong,et al.Estimate of X-radiation to patients during transhepatic artery chemoembolization for heptocarcinoma[J].Chinese Journal of Radiological Medicine and Protection,2004,24(5):472-474
Estimate of X-radiation to patients during transhepatic artery chemoembolization for heptocarcinoma
Received:September 23, 2003  
DOI:
KeyWords:X-ray radiation  Hepatocarcinoma  Chemoembolization
FundProject:
Author NameAffiliationE-mail
CHEN Sheng-li Department of Radiology, The First Municipal People's Hospital of Guangzhou 510180, China csl@cha.163.com 
ZHU Dong-liang Department of Radiology, The First Municipal People's Hospital of Guangzhou 510180, China  
CHEN Guo-dong Department of Radiology, The First Municipal People's Hospital of Guangzhou 510180, China  
邹蓉珠 广州市疾病预防控制中心  
黄齐好 Department of Radiology, The First Municipal People's Hospital of Guangzhou 510180, China  
黄子诚 Department of Radiology, The First Municipal People's Hospital of Guangzhou 510180, China  
谢琦 Department of Radiology, The First Municipal People's Hospital of Guangzhou 510180, China  
慕容爱 Department of Radiology, The First Municipal People's Hospital of Guangzhou 510180, China  
Hits: 4705
Download times: 2263
Abstract::
      Objective To evaluate the X- radiation to patients of hepatocarcinoma during transhepatic artery chemoembolization(THACE). Methods The radiation dose was measured in 82 patients(group A) during THACE for hepatocarcinoma.Measurements were carried out using a dosimeter system(Diamentor K1 and Diamentor ED) equipped in DSA setting Angiostar-Plus(Siemens,Germany),dose-area product(DAP) and entrance surface dose(ESD) were recorded on-line,and the effect dose(ED) was estimated by Monte-Carlo conversion coefficient from DAP.The effects of elevated tube voltage,reduced pulsed fluoroscopy(PF) frequency and curtailed total photography frames on radiation to the patients were studied by comparison with another 10 THACE cases treated recently(group B). Results In group A,the mean fluoroscopic time was (35.3±21.1) minutes,and the mean number of angiographic exposure was (234±108) frames,the mean DAP dose was((21 748±)(12 424)) cGy/cm2,and the mean ESD was (964±632)mGy,ED was (34.8±19.9)mSv.The contribution of pulsed fluoroscopy(PF) dose to the total DAP(24.0±12.7)% was smaller than photography(R)(75.9±10.7)%,and of PF to total ESD(49.8±14.9)% was similar to R (51.7±14.2)%.The total dose,fluoroscopy dose per minute and photography dose per frame in group B was reduced obviously than in group A. Conclusion The HCC patients suffered a large dose of X-ray exposure in THACE procedure.The radiation exposure of patients can be reduced by elevating tube voltage,reducing pulsed fluoroscopy frequency and curtailing total photography frames moderately.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12035931  On-line:0

v
Scan QR Code
&et=C62E7D20909F40ABD6289A95A8C4A9C579A3994BF33BF547C1951C310C116EE59183F029CBEED703576D1F05A7584E4D6F12A7FDFE131AA2751B3F64DA23B4DC5B2CAD9C3008ABB38DE82DCCC0188C8D939EC4C5E6A792403314F778949FCFD21F32F3520B080DBE&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=D0E58B75BFD8E51C&aid=D38B0BE1D2EB2F78&vid=&iid=94C357A881DFC066&sid=7D1E6EEC2019967D&eid=FE6B7E9BDCCDBAA6&fileno=20040540&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="D0E58B75BFD8E51C"; var my_aid="D38B0BE1D2EB2F78";