CHENG Jin sheng,LI Kai bao,ZHAO Zhao luo,et al.The research of model RT-101 thimble ionization chamber[J].Chinese Journal of Radiological Medicine and Protection,2004,24(3):275-277 |
The research of model RT-101 thimble ionization chamber |
Received:March 02, 2004 |
DOI: |
KeyWords:Radiotherapy Absorbed dose Ionization chamber Energy response |
FundProject:卫生部医学重点科技基金资助项目(88113069) |
Author Name | Affiliation | CHENG Jin sheng | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | LI Kai bao | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | ZHAO Zhao luo | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China | 姜庆寰 | National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China |
|
Hits: 3763 |
Download times: 2508 |
Abstract:: |
Objective To research the thimble ionization chamber for measurement of absorbed dose in photon or electron radiation field used in radiotherapy. Methods Using graphite as chamber wall and pure aluminum as central receiver electrode. Result According to the IEC standard 60731 checking methods, all properties of RT-101 are up to the requirements of the standard. Conclusion The 0.6 cm3 RT-101 thimble ionization chamber can be used as field class detector for measurement of absorbed dose in photon or electron radiation field used in radiotherapy. |