JIA Mingxuan,ZOU Huawei,WU Rong,et al.Measurement and analysis of the thoracic patient setup deviations in routine radiotherapy[J].Chinese Journal of Radiological Medicine and Protection,2003,23(3):193-195
Measurement and analysis of the thoracic patient setup deviations in routine radiotherapy
Received:December 25, 2002  
DOI:
KeyWords:Thoracic irradiation  EPID  Setup deviations
FundProject:辽宁省自然科学基金资助项目(20022070)
Author NameAffiliation
JIA Mingxuan Center of Radiation Oncology, 2nd Hospital, China Medical University, Shengang 110004, China 
ZOU Huawei Center of Radiation Oncology, 2nd Hospital, China Medical University, Shengang 110004, China 
WU Rong Center of Radiation Oncology, 2nd Hospital, China Medical University, Shengang 110004, China 
孙健 Center of Radiation Oncology, 2nd Hospital, China Medical University, Shengang 110004, China 
董晓棋 Center of Radiation Oncology, 2nd Hospital, China Medical University, Shengang 110004, China 
Hits: 3654
Download times: 2312
Abstract::
      Objective To determine the magnitude of the setup deviations of the thoracic patients in routine radiotherapy. Methods Altogether 408 films for 21 thoracic patients were recorded using the electronic portal imaging device (EPID), and comparison with reference CT simulator digitally-reconstructed radiograph (DRR) for anterior-posterior fields was performed.The deviation of setup for 21 patients in the left-right (RL), superior-inferior(SI)directions and rotation about the anterior-posterior (AP) axis were measured and analyzed. Results Without immobilization device, the mean translational and rotational setup deviations were (0. 7±3. 1) mm and (1.5±4. 1) mm in the RL and SI directions, respectively, and (0. 3±2. 4)° about AP axis.With immobilization device, the mean translational and rotational setup deviations were (0.5±2.4) mm and (0.8±2. 7) mm in the RL and SI directions respectively, and (0.2±1.6)° about AP axis. Conclusion The setup deviations in thoracic patients, irradiation may be reduced with the use of the immobilization device.The setup deviation in the SI direction is greater than that in the RL direction.The setup deviations are mainly random errors.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:11567953  On-line:0

v
Scan QR Code
&et=DC44477595533BA79D4991F2B6CA64D03C500A3816E786482C61D3F4C761F4D3A5A34F06B294EF8D59800D720BD1C6C5586B9A84F562A51D2E534DA908CC081653D9BCDFF27570BE17FEBED111DE19AD477E73386DA7A55BD6C3E14C37E52FFF&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=D43C4A19B2EE3C0A&aid=27D873A3723F16AE&vid=&iid=38B194292C032A66&sid=23104246A5FCFCEF&eid=64963996248CBF47&fileno=20030321&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="D43C4A19B2EE3C0A"; var my_aid="27D873A3723F16AE";