CHENG Jinsheng,GUO Zhaohui,AN Jinggang.Development of RTS-200S 2D automatic scanning dose measuring water phantom for radiotherapy[J].Chinese Journal of Radiological Medicine and Protection,2003,23(3):190-192
Development of RTS-200S 2D automatic scanning dose measuring water phantom for radiotherapy
Received:December 28, 2002  
DOI:
KeyWords:D Dose  Scanning measurement  Radiotherapy
FundProject:
Author NameAffiliation
CHENG Jinsheng National Institute for Rodiological Protection and Nuclear Safety, China CDC, Beijing 100088, China 
GUO Zhaohui National Institute for Rodiological Protection and Nuclear Safety, China CDC, Beijing 100088, China 
AN Jinggang National Institute for Rodiological Protection and Nuclear Safety, China CDC, Beijing 100088, China 
姜庆寰 National Institute for Rodiological Protection and Nuclear Safety, China CDC, Beijing 100088, China 
刘立明 National Institute for Rodiological Protection and Nuclear Safety, China CDC, Beijing 100088, China 
李延波 National Institute for Rodiological Protection and Nuclear Safety, China CDC, Beijing 100088, China 
李开宝 National Institute for Rodiological Protection and Nuclear Safety, China CDC, Beijing 100088, China 
Hits: 2593
Download times: 2018
Abstract::
      Objective To developt a 2D automatic scanning water phantom, being applicable for acceptance tests of radiotherapy equipment and quick measurements of clinical physical parameters for radiotherapy beams and suitable for use in radiotherapy departments and provincial monitoring institutions. Methods Two chambers were used as a scanning detector and a beam monitor, respectively.Modern automatic techniques were adopted in the scanning system so as to realize rapid and accurate measurement for medical linear accelerator, X-knife and Co-60 therapy beams.The precision of scanning position was less than±0. 1 mm by means of computer-controlled stepper motors. Results RTS-200S enabled user to acquire data of beam profiles, depth dose curve.RTP software program integrated data collec`tion, analysis, output and storage.The measuring results were in agreement with those measured by Varian company. Conclusion The RTS-200S water phantom and electronic control are excellent in both accuracy and reliability for measuring beam speciality of radiotherapy units.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:9069419  On-line:0

v
Scan QR Code
&et=A90EAC7419FB9B224D2711078F315DBFC82D815746B6567C8E9814D23FE786498F089F01FB2D77BE0FCA1818FE7BCAAFCFDC66E84A1D4380345FD738D19E22A8E725F4AA2CF9EE289A153FF8A416BE65D8634D921914A07F9BE67E845E52B6F9&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=D43C4A19B2EE3C0A&aid=3FA6195ACED8F1D7&vid=&iid=38B194292C032A66&sid=6235172E4DDBA109&eid=C29816B2656377A7&fileno=20030320&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="D43C4A19B2EE3C0A"; var my_aid="3FA6195ACED8F1D7";