XU Zhiyong,ZHANG Liang-an,DAI Guangfu,et al.Dose distribution calculated by Monte Carlo method in a β-coated stent[J].Chinese Journal of Radiological Medicine and Protection,2002,22(6):431-433
Dose distribution calculated by Monte Carlo method in a β-coated stent
Received:November 16, 2001  
DOI:
KeyWords:Monte Carlo method  Radioactive stent  Restenosis  Dosimetry
FundProject:国家自然科学基金资助项目:(30170233)
Author NameAffiliation
XU Zhiyong Intitute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medicine College, Tianjin 300192, China 
ZHANG Liang-an Intitute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medicine College, Tianjin 300192, China 
DAI Guangfu Intitute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medicine College, Tianjin 300192, China 
苑淑渝 Intitute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medicine College, Tianjin 300192, China 
范亚光 Intitute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medicine College, Tianjin 300192, China 
李运芳 Intitute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medicine College, Tianjin 300192, China 
杨翊 Intitute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medicine College, Tianjin 300192, China 
Hits: 3771
Download times: 2148
Abstract::
      Objective To compare the results of distribution of dose rate calculated by numerical integration and the Monte Carlo methods. Methods The distribution of dose fate of the stent was estimated with three typical dose-point-kernel functions. Results The radial dose distribution(CDD) on center surface,axial DDON the surface and 0.5 mm apart from the surface of the stent were calculated.For the radial distribution,the variation was less than 1.5%,and so the axial one. Conclusion The results of dose distribution calculated with two methods very close, so the Monte Carlo method can be applied for calculation the dose distribution of radioactive stent. ;
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:12029977  On-line:0

v
Scan QR Code
&et=376AE8D779458D135AF998DF1828505A4D63240157CAB9135C292EECDBEFB2DA0417DFDBEDEAE29240D23289E2B218D11A2307674E6D967A283868D762D764CF7F9352B2BA9FFF0A&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=C3ACC247184A22C1&aid=74A1AF5F6916D02B&vid=&iid=B31275AF3241DB2D&sid=6D6BFCF0101BC091&eid=DDEED1BDDBFAA8A7&fileno=20020619&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="C3ACC247184A22C1"; var my_aid="74A1AF5F6916D02B";