Ji Gang,Guo Yong,Zhu Yunping,et al.Measurement and calculation of tissue air ratios of cobalt-60 gamma radiation in large field and depth[J].Chinese Journal of Radiological Medicine and Protection,1997,17(3):169-172
Measurement and calculation of tissue air ratios of cobalt-60 gamma radiation in large field and depth
Received:November 20, 1996  Revised:February 20, 1997
DOI:
KeyWords:Tissue air ratio  Absorbed dose  Phantom  Depth  Field  60Co γ-radiation
FundProject:
Author NameAffiliation
Ji Gang Institute of Radiation Medicine, Beijing 100850, China 
Guo Yong Institute of Radiation Medicine, Beijing 100850, China 
Zhu Yunping Institute of Radiation Medicine, Beijing 100850, China 
周振山 Institute of Radiation Medicine, Beijing 100850, China 
王兴功 Institute of Radiation Medicine, Beijing 100850, China 
杨甲祥 Institute of Radiation Medicine, Beijing 100850, China 
Hits: 5005
Download times: 1993
Abstract::
      Objective Tissue air ratio(TAR) is the one of useful concepts and methods in radiation dosimetry.There are no available TAR data for large field and depth irradiation and more research is needed to investigate the properties of TAR and other relative quantities under special irradiation. Methods Data of TAR for depth up to 70.0cm and rectangular field size up to 50.0cm×50.0cm have been obtained for 60Coγ-rays. Results For depth less than 30.0cm and field size less than 50.0cm×50.cm, the relative errors between the data by the experiment and by reference are less than 5 percent. Conclusions It is very important to measure TAR under the condition of irradiation in large depth and field sizes for the measurment and calculation of dose distribution in human body after a radiological accident.
HTML  View Full Text  View/Add Comment  Download reader
Close

Copyright©    Editorial Office of Chinese Journal of Radiological Medicine and Protection    

Beijing ICP No. 05020547 -2

Address: 2 Xinkang Street, Dewai, Beijing 100088, China

Telephone:010-62389620; Email:cjrmp@cjrmp.sina.net

Technical Support:Beijing E-tiller CO.,LTD.

Visitors:17874464  On-line:0

v
Scan QR Code
&et=DC44477595533BA79D4991F2B6CA64D0656F82443B4C7EAC6FBE2D8026DF824DAD487C8FE6FD0538E876C210C9DB34E8F2BF4CF5CC4440770E8FA7280A5475AAD09DD26BCCA253AE12679F9098C5887F8AD7CF8E8CC7D1661E566A8DC3BEB4A3BBD8BF706E02C246&pcid=A9DB1C13C87CE289EA38239A9433C9DC&cid=D4D466D60FDC1A5A&jid=5E4353813E091AB841B02B880782B82C&yid=5370399DC954B911&aid=43B1C4D406DE3C34221F804E0C4AA632&vid=&iid=38B194292C032A66&sid=954CE65414DD94CA&eid=9D453329DCCABB94&fileno=19970312&flag=1&is_more=0"> var my_pcid="A9DB1C13C87CE289EA38239A9433C9DC"; var my_cid="D4D466D60FDC1A5A"; var my_jid="5E4353813E091AB841B02B880782B82C"; var my_yid="5370399DC954B911"; var my_aid="43B1C4D406DE3C34221F804E0C4AA632";