Yuan Shilin,Dai Hongchen,Wu Dake.Optimization of Measurement Conditions otf Low-Level TLD Systems[J].Chinese Journal of Radiological Medicine and Protection,1985,5(1):34-37 |
Optimization of Measurement Conditions otf Low-Level TLD Systems |
Received:January 03, 1983 |
DOI: |
KeyWords:TLD system quality factor optimization |
FundProject: |
Author Name | Affiliation | Yuan Shilin | Institute of Radiomedicine, Chinese Academy of Medical Sciences, Tianjin | Dai Hongchen | Institute of Radiomedicine, Chinese Academy of Medical Sciences, Tianjin | Wu Dake | Institute of Radiomedicine, Chinese Academy of Medical Sciences, Tianjin | 孙福印 | Institute of Radiomedicine, Chinese Academy of Medical Sciences, Tianjin | 杨和理 | Institute of Radiomedicine, Chinese Academy of Medical Sciences, Tianjin | 苑淑瑜 | Institute of Radiomedicine, Chinese Academy of Medical Sciences, Tianjin |
|
Hits: 2370 |
Download times: 1665 |
Abstract:: |
The quality factor P=S0/B of thermoluminescent dosemeter system was deduced in thiB paper.In the case of LiF-HS and JRG Ⅲ TLD system, when minimal values in the objective function P=f(T, T1, T2…)were chosen, a set of relevant measure parameters was obtained in the experimental method of optimization. This made the applicability of the system extend to low-dose level for aiq dcder of magnitude.When this method was applied to JR-152, FJ-337 TLD systemspthe results were also obviOUS.It can be used.In any other TLD systems.The quality fadtot can be used as a criterion for checking TLD system and for comparing different TLD systems. |