Wu Dake,Shun Fuyin,Dai Hongchen.A HIGHLY SENSITIVE LITHIUM FLUORIDE THERMOLUMINESCENCE DOSIMETER[J].Chinese Journal of Radiological Medicine and Protection,1981,1(5):13-19 |
A HIGHLY SENSITIVE LITHIUM FLUORIDE THERMOLUMINESCENCE DOSIMETER |
|
DOI: |
KeyWords: |
FundProject: |
|
Hits: 2914 |
Download times: 2250 |
Abstract:: |
A New highly sensitive LiF(Mg,Cu,P)has been made in our laboratory.Its dosimetric properties have been studied.The main results obtained in this experiment are as follows:The sensitivity of new highly sensitive LiF TLD is about 30 times higher than that of Harshaw LiF TLD-100。And nearly equal to that of CaSO4(TLD;the thermoluminescence response to the exposure increases linearly in the range of the exposure from 0.3 mR to 1000R;the energy dependence of LiF(Mg,CU, P)TLD is so small that its response at 30 keV is 1.26 times as large as that to 60Co gamma radiation;its fading approximates to 5% at room temperature during two months. |
HTML View Full Text View/Add Comment Download reader |