陶莉,张慧,吴翊凯,等.浅层X射线治疗机过滤器对放疗剂量学影响的蒙特卡罗模拟研究[J].中华放射医学与防护杂志,2025,45(3):194-201.Tao Li,Zhang Hui,Wu Yikai,et al.Monte Carlo simulation study of the effect of filter on radiotherapy dosimetry in superficial X-ray therapy apparatus[J].Chin J Radiol Med Prot,2025,45(3):194-201
浅层X射线治疗机过滤器对放疗剂量学影响的蒙特卡罗模拟研究
Monte Carlo simulation study of the effect of filter on radiotherapy dosimetry in superficial X-ray therapy apparatus
投稿时间:2023-07-27  
DOI:10.3760/cma.j.cn112271-20240327-00109
中文关键词:  浅层X射线放疗  过滤器  蒙特卡罗模拟  剂量优化
英文关键词:Superficial X-ray radiotherapy  Filter  Monte Carlo simulation  Dosimetric optimization
基金项目:安徽省重点研究与开发计划(2023s07020020);国家自然科学基金项目(12275372);中国科学技术大学双一流医学物理与生物医学工程交叉学科平台(YD2140000601);2020年中科大新医学创新团队项目(YD2140002002)
作者单位E-mail
陶莉 中国科学技术大学核科学技术学院, 合肥 230025  
张慧 中国科学技术大学核科学技术学院, 合肥 230025
中国科学技术大学附属第一医院放疗科, 合肥 230031 
 
吴翊凯 中国科学技术大学核科学技术学院, 合肥 230025  
刘君怡 中国科学技术大学核科学技术学院, 合肥 230025  
齐妙 中国科学技术大学核科学技术学院, 合肥 230025  
高宁 中国科学技术大学核科学技术学院, 合肥 230025  
常艳奎 中国科学技术大学核科学技术学院, 合肥 230025  
裴曦 安徽慧软科技有限公司, 合肥 230088  
陈志 中国科学技术大学核科学技术学院, 合肥 230025  
徐榭 中国科学技术大学核科学技术学院, 合肥 230025 xgxu@ustc.edu.cn 
摘要点击次数: 1622
全文下载次数: 750
中文摘要:
      目的 探索针对使用X射线管的浅层放射治疗机基于过滤器厚度和形状选择的剂量学优化策略。方法 使用蒙特卡罗代码TOPAS对X射线管配套设备建模,在水模体中模拟6种X射线能量(50~150 kV)以及5种不同厚度的常规铝过滤器(0.5~3.0 mm)的剂量分布,分析了沿中心轴的百分比深度剂量(PDD)曲线、不同深度的中心轴剖面二维剂量以及横向剂量曲线。同时,对比了3种不同设计的铝制过滤器(常规圆柱形、圆锥形和斜柱形过滤器)的剂量分布,以评估不同过滤器形状的剂量学优化的效果。结果 在同一能量下,增加过滤器厚度可以优化浅表皮肤剂量,70 kV能量下深度剂量均匀性优化效果在5.5 mm深处最大能提高26%。应用常规圆柱形、锥形和斜柱形的铝过滤器,能够分别实现凸起、平坦和倾斜的剂量分布模式。结论 通过合理选择X射线能量和过滤器厚度,可以获得与靶区深度相匹配的理想剂量分布;引入个性化过滤器对于多样性靶区目标也具有重要意义。
英文摘要:
      Objective To explore the dosimetry optimization strategy based on filter thickness and shape selection for the bulb superficial X-ray radiotherapy unit. Methods Monte Carlo code TOPAS was used to model tubular equipment, and the dose distribution from six X-ray energies (50-150 kV) and five conventional aluminum filters (0.5-3.0 mm) with different thickness were simulated in the water model. The percentage depth dose (PDD) curve along the central axis, the center-axis profile dose at different depths, and the lateral dose distribution were analyzed. The dose distribution of three different designs of aluminum filters (conventional cylindrical, conical and oblique cylindrical filters) was compared to evaluate the effect of dosimetric optimization of different filter shapes. Results Under the same energy, increasing the thickness of the filter can optimize the superficial skin dose, and the optimization effect of depth dose uniformity can be increased by 26% at a depth of 5.5 mm at 70 kV energy. The raised, flat and inclined dose distribution modes can be achieved by using conventional cylindrical, conical and inclined aluminum filters. Conclusions By selecting the appropriate X-ray energy and filter thickness, an ideal dose distribution matching the tumor depth can be achieved. The application of personalized filters is also of great significance for diverse target areas.