赵强,吴湘阳,常晓斌,冯涛,杨迪,张坤,屈喜梅,王学敏,邓佳.二维矩阵电离室验证Acuros BV算法计算相对剂量分布的可行性研究[J].中华放射医学与防护杂志,2020,40(1):59-63
二维矩阵电离室验证Acuros BV算法计算相对剂量分布的可行性研究
Feasibility study of using two dimensional array ion-chamber to verifiy relative dose distribution calculated with Acuros BV
投稿时间:2019-05-16  
DOI:10.3760/cma.j.issn.0254-5098.2020.01.011
中文关键词:  二维矩阵电离室  Acuros BV  近距离治疗
英文关键词:2D array ion-chamber  Acuros BV  Brachy therapy
基金项目:
作者单位E-mail
赵强 陕西省肿瘤医院放疗科, 西安 710061  
吴湘阳 陕西省肿瘤医院放疗科, 西安 710061 13087566100@163.com 
常晓斌 陕西省肿瘤医院放疗科, 西安 710061  
冯涛 陕西省肿瘤医院放疗科, 西安 710061  
杨迪 陕西省肿瘤医院放疗科, 西安 710061  
张坤 陕西省肿瘤医院放疗科, 西安 710061  
屈喜梅 陕西省肿瘤医院放疗科, 西安 710061  
王学敏 陕西省肿瘤医院放疗科, 西安 710061  
邓佳 陕西省肿瘤医院放疗科, 西安 710061  
摘要点击次数: 2072
全文下载次数: 1227
中文摘要:
      目的 研究PTW729二维矩阵电离室验证Acuros BV算法在相对剂量分布计算方法的可行性,分析该方法的优势和劣势,为实际临床工作提供参考。方法 基于自建测量模体,在相同计划和测量条件下,使用PTW729和胶片方法,分别测试相同位置处的剂量分布,对两种方法的测量结果和Acuros BV算法计算结果分别进行γ比对分析,并测试PTW729方法的稳定性。结果 PTW729和胶片的测试结果与Acuros BV算法计算结果的γ比对值分别为98.9%、95.9%,PTW729与胶片测试结果的γ比对值为88.0%。PTW729稳定性测试实验γ比对值分别为95.0%、100.0%、100.0%。结论 PTW729二维矩阵电离室可应用于Acuros BV算法相对剂量分布的快速验证。
英文摘要:
      Objective To study the feasibility of using the PTW729 2D array ion-chamber to verify the relative dose distribution calculated with the Acuros BV algorithm. Both advantages and disadvantages of the method were analyzed to provide reference for practical clinical practices. Methods Based on self-built measurement phantoms, the dose distribution on the same slice of the phantom was measured with PTW729 and film, respectively, under the same measurement condition and plan. The dose distributions obtained by the two method were compared with the result calculated with Acuros BV, separately, by using γ analytical tool. And the stability of the PTW729 was tested. Results The γ comparison value was 95.9% between the film and Acuros BV, 98.9% between the PTW729 and Acuros BV and 88.0% between the film and PTW729, with 95.0%, 100.0%, and 100.0%, in their stability test respectively. Conclusions PTW729 2D array ion-chamber can be applied to the rapid verification of Acuros BV algorithm-calculated relative dose distribution.