陈掌凡,赵新春,张会敏,等.广西X射线诊断受检者体表入射剂量调查[J].中华放射医学与防护杂志,2016,36(4):298-302.Chen Zhangfan,Zhao Xinchun,Zhang Huimin,et al.Investigation on entrance surface doses to patients from medical X-ray diagnosis in Guangxi[J].Chin J Radiol Med Prot,2016,36(4):298-302 |
广西X射线诊断受检者体表入射剂量调查 |
Investigation on entrance surface doses to patients from medical X-ray diagnosis in Guangxi |
投稿时间:2015-08-04 |
DOI:10.3760/cma.j.issn.0254-5098.2016.04.014 |
中文关键词: X射线诊断 体表入射剂量 直接数字化X射线摄影 计算机X射线摄影 医疗照射 |
英文关键词:X-ray diagnosis Entrance surface dose Digital radiography Computed radiography Medical exposure |
基金项目:卫生行业科研专项(201002009);广西自然科学基金(2014GXNSFAA118223) |
作者 | 单位 | E-mail | 陈掌凡 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 赵新春 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 张会敏 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 唐孟俭 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 梁挺 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 陈发想 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 董颖 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 马一龙 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 覃志英 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | | 谢萍 | 530028 南宁, 广西壮族自治区疾病预防控制中心放射卫生防护所 | gxcdcxp@126.com |
|
摘要点击次数: 5309 |
全文下载次数: 2884 |
中文摘要: |
目的 探讨广西X射线诊断受检者体表入射剂量。方法 分层随机抽样选取25家不同级别医院的2 236例X射线诊断受检者为调查对象,采用热释光剂量测量不同级别的医院、不同类型设备和不同照射部位受检者的体表入射剂量。结果 普通X射线摄影、计算机X射线摄影(CR)和直接数字化X射线摄影(DR)受检者体表入射剂量范围分别为0.08~31.51、0.11~4.25和0.05~35.63 mGy。腹部前后位(AP)、骨盆AP;头颅侧位(LAT)、头颅后前位(PA)、胸部PA、胸部LAT、胸椎AP、胸椎LAT、腰椎AP、腰椎LAT入射剂量范围分别为0.08~19.53、0.15~18.78、0.08~9.87、0.06~9.24、0.05~2.71、0.13~2.93、0.15~19.01、0.07~25.33、0.16~27.23和0.11~35.63 mGy。结论 广西X射线诊断受检者平均入射剂量达标,但部分DR摄影致胸部PA入射剂量超过医疗照射指导水平。 |
英文摘要: |
Objective To investigate the entrance surface doses to the patients undergoing diagnostic X-ray examinations in Guangxi. Methods A total of 2236 patients who have received diagnostic X-ray examinations were investigated at 25 randomly selected hospitals at different levels where dthermoluminescence dosimeters were used to measure the entrance surface doses in different hospitals with different equipment and different positions. Results The entrance surface dose from X-ray imaging, computed radiography (CR) and digital radiography (DR) were in the range of 0.08-31.51, 0.11-4.25 and 0.05-35.63 mGy, respectively. The entrance doses to abdomen (AP), pelvis (AP), head (LAT), head (PA), chest (PA), chest (LAT), thoracic vertebra (AP), thoracic vertebra (LAT), lumbar vertebra (AP), lumbar vertebra (LAT) were 0.08-19.53, 0.15-18.78, 0.08-9.87, 0.06-9.24, 0.05-2.71, 0.13-2.93, 0.15-19.01, 0.07-25.33, 0.16-27.23, and 0.11-35.63 mGy, respectively. Conclusions The average dose to patients from X-ray diagnosis in Guangxi was in accordance with the relevant standards, but DR could partly lead the chest (PA) dose to exceed the guidance levels of medical exposure. |