于路新,何奕松,殷云鹏,等.宫颈癌后装近距离放疗中TG43算法与AcurosBV算法的剂量学比较[J].中华放射医学与防护杂志,2026,46(7):670-675.Yu Luxin,He Yisong,Yin Yunpeng,et al.Dosimetric comparison between TG43 and AcurosBV algorithms in brachytherapy for cervical cancer[J].Chin J Radiol Med Prot,2026,46(7):670-675
宫颈癌后装近距离放疗中TG43算法与AcurosBV算法的剂量学比较
Dosimetric comparison between TG43 and AcurosBV algorithms in brachytherapy for cervical cancer
投稿时间:2025-09-03  
DOI:10.3760/cma.j.cn112271-20250903-00314
中文关键词:  宫颈癌  后装近距离放疗  TG43算法  AcurosBV算法  剂量学比较
英文关键词:Cervical cancer  Brachytherapy  TG43 algorithm  AcurosBV algorithm  Dosimetric comparison
基金项目:
作者单位E-mail
于路新 四川省医学科学院 四川省人民医院肿瘤中心, 成都 610072  
何奕松 四川省医学科学院 四川省人民医院肿瘤中心, 成都 610072 627412679@qq.com 
殷云鹏 四川省医学科学院 四川省人民医院肿瘤中心, 成都 610072  
张薇莎 四川省医学科学院 四川省人民医院肿瘤中心, 成都 610072  
梁田甜 四川省医学科学院 四川省人民医院肿瘤中心, 成都 610072  
骆科林 四川省医学科学院 四川省人民医院肿瘤中心, 成都 610072  
摘要点击次数: 1243
全文下载次数: 5
中文摘要:
      目的 比较宫颈癌后装放疗中使用不锈钢曼彻斯特施源器时TG43算法与AcurosBV算法的剂量学差异,为临床精准剂量计算提供依据。方法 回顾性分析四川省人民医院21例使用不锈钢曼彻斯特施源器的宫颈癌患者,在瓦里安Eclipse计划系统中分别采用TG43算法和AcurosBV算法计算剂量,统计高危临床靶区(HRCTV)D90%D98%及膀胱、直肠、小肠、尿道等危及器官(OAR)D2 cm3D0.1 cm3剂量差异。开发基于Python的高密度体素自动识别工具,提升AcurosBV算法的计算成功率。结果 AcurosBV算法的HRCTV剂量显著低于TG43算法,D90%降低了7.46%(t=91.88,P<0.01),D98%降低了7.27%(t=39.47,P<0.01);危及器官剂量降幅为5.53%~8.80%(P<0.01),其中小肠D2 cm3差异最大(-8.80%)。Python工具可有效解决临床漏检问题。结论 AcurosBV算法在不锈钢施源器中表现出显著的剂量修正效应与更优的计算精度。本研究结果为TG43算法向模型化剂量计算算法(MBDCA)的临床转化提供了必要的剂量学参考依据。
英文摘要:
      Objective To evaluate dosimetric differences between TG43 and AcurosBV algorithms for stainless-steel Manchester applicators in cervical cancer brachytherapy, and to provide a basis for precise clinical dose calculation.Methods A retrospective study analyzed 21 patients with cervical cancer who were treated with stainless-steel Manchester applicators at Sichuan Provincial People’s Hospital. Doses were calculated using TG43 and AcurosBV algorithms in Varian Eclipse, with comparisons of D90% and D98% of high-risk clinical target volume (HRCTV) as well as D2 cm3 and D0.1 cm3 of (P<0.01) organs at risk (OARs), including the bladder, rectum, small intestine, and urethra. A Python-based automated tool for high-density voxel identification was developed to improve the computational success rate of the AcurosBV algorithm. Results AcurosBV yielded significantly lower doses than TG43 for HRCTV, with 7.46% and 7.27% reductions for D90% and D98%, respectively(t=91.88,39.47,P<0.01); the decreases in doses for OARs ranged from 5.53% to 8.80%(P<0.01), with the largest difference in small intestine D2 cm3(-8.80%). The Python tool effectively addressed clinical under-detection. Conclusions For stainless-steel applicators, the AcurosBV algorithm demonstrated a significant dose correction effect and superior calculation accuracy. These results provide an essential dosimetric reference for the clinical transition from TG43 to model-based dose calculation algorithms.