郑仁超,杨想军,卢晓光,等.国内首台单间室质子治疗系统场所辐射水平的测量与评估[J].中华放射医学与防护杂志,2024,44(10):879-885.Zheng Renchao,Yang Xiangjun,Lu Xiaoguang,et al.Measurements and assessment of radiation levels at the radiotherapy site of the first domestic single-vault proton therapy system[J].Chin J Radiol Med Prot,2024,44(10):879-885
国内首台单间室质子治疗系统场所辐射水平的测量与评估
Measurements and assessment of radiation levels at the radiotherapy site of the first domestic single-vault proton therapy system
投稿时间:2024-02-07  
DOI:10.3760/cma.j.cn112271-20240207-00052
中文关键词:  单间室  质子  治疗场所  辐射水平
英文关键词:Single-vault  Proton  Radiotherapy place  Radiation level
基金项目:湖北省科技计划项目(2023BCB096)
作者单位E-mail
郑仁超 华中科技大学同济医学院附属同济医院肿瘤科, 武汉 430030  
杨想军 湖北省疾病预防控制中心, 武汉 430079  
卢晓光 深圳清华大学研究院精准放射诊疗技术研发中心, 深圳 518063 luxg@tsinghua-sz.org 
吴韦清 华中科技大学同济医学院附属同济医院肿瘤科, 武汉 430030  
彭知义 华中科技大学同济医学院附属同济医院肿瘤科, 武汉 430030  
胡广原 华中科技大学同济医学院附属同济医院肿瘤科, 武汉 430030  
袁响林 华中科技大学同济医学院附属同济医院肿瘤科, 武汉 430030  
摘要点击次数: 1569
全文下载次数: 518
中文摘要:
      目的 测量并评估国内首台单间室质子治疗系统场所内的相关辐射剂量。方法 测量治疗系统出束时和停束后治疗场所的辐射水平,并评估场所内人员年有效剂量。结果 出束过程中,探测到最高辐射剂量在设备层防护门处,γ辐射水平为2.140 μSv/h,中子辐射水平为0.850 μSv/h。束流停止后,中子辐射随即消失,活化产物的辐射以γ射线为主,同一地点停束时间越长,感生辐射强度越低。同一时间与受照物体的距离越远,感生辐射强度越低。评估所有人员受照年有效剂量均处于安全水平,其中物理师最高,为2.138 mSv。结论 本中心质子治疗场所辐射水平符合要求,可以安全开展质子治疗工作。
英文摘要:
      Objective To measure and assess relevant radiation doses at the radiotherapy site of the first domestic single-vault proton therapy system. Methods The radiation levels of the therapy system during and after beam irradiation were measured, and annual effective doses were assessed for personnel at the site. Results During beam irradiation, the highest radiation dose was detected at the shielded door of the equipment floor, with a gamma radiation level of 2.140 μSv/h and a neutron radiation level of 0.850 μSv/h. Neutron radiation disappeared immediately once the beams stopped. In contrast, the radiation activated originated mostly from gamma rays. A longer time after beams stopped was associated with lower induced radiation intensity at the same location. Furthermore, a farther distance from the irradiated object corresponded to lower induced radiation intensity at the same time. The assessment result reveal that the annual effective doses to the personnel were at the safe level, with physicists exposed to the highest dose of 2.138 mSv. Conclusions The radiation level at the studied proton therapy site meets the safety requirement, and the treatment can be performed safely at this site.