张坤,石祥,谢双双,等.利用同源双能CT (SF-DECT)提高儿童腹部图像质量的可行性研究[J].中华放射医学与防护杂志,2023,43(8):639-644.Zhang Kun,Shi Xiang,Xie Shuangshuang,et al.Feasibility of improving the quality of children's abdominal images using split-filter dual-energy CT[J].Chin J Radiol Med Prot,2023,43(8):639-644 |
利用同源双能CT (SF-DECT)提高儿童腹部图像质量的可行性研究 |
Feasibility of improving the quality of children's abdominal images using split-filter dual-energy CT |
投稿时间:2023-02-13 |
DOI:10.3760/cma.j.cn112271-20230213-00034 |
中文关键词: 双能CT 儿童体模 辐射剂量 图像质量 |
英文关键词:Dual-energy CT Child phantom Radiation dose Image quality |
基金项目:国家自然科学基金(81901710,81873888);天津市卫生健康科技项目青年人才项目(TJWJ2021QN011,QN20024);天津市自然科学基金(22JCQNJC00920);天津市医学重点学科(专科)建设项目(TJYXZDXK-041A)临床试验注册:中国临床试验注册中心,ChiCTR1800019231 |
|
摘要点击次数: 2103 |
全文下载次数: 906 |
中文摘要: |
目的 评估在儿童腹部检查中,同源双能CT (SF-DECT)是否能在低剂量条件下提高图像质量。方法 采用儿童体模进行初步研究,并于2020年6月至12月期间前瞻性招募20名4~6岁的儿童进行临床验证。分别采用常规单能CT (SECT)和SF-DECT扫描体模和儿童的的腹部区域,比较SF-DECT和SECT在各剂量下(1、2、3和4 mGy) CT值、图像噪声、对比噪声比(CNR)和图像主观评分。结果 在体模研究中,3和4 mGy剂量条件下,与SECT相比,SF-DECT的图像噪声分别降低了18.9%和23.6%,肝脏和肾脏CNR分别最多增加了12.8%和31.9%(Z=3.00、5.17,P<0.001)。在儿童SF-DECT扫描中,随着扫描剂量的增加,图像噪声降低(Z=4.64,P<0.001),肝脏和肾脏CNR增加(Z=3.78、3.39,P<0.001)。在体模研究和患者研究中,SF-DECT扫描图像的主观评分均优于SECT (Z=1.96~3.80,P<0.05)。结论 与SECT相比,SF-DECT能够提高儿童腹部扫描图像质量。该技术在优化儿童腹部CT扫描方面有一定前景,但需要进行深入的临床研究来验证结果。 |
英文摘要: |
Objective To evaluate the effects of split-filter dual-energy CT (SF-DECT) in improving image quality at low doses in the process of abdominal examinations for children.Methods A preliminary study was conducted using child phantoms. Furthermore, 20 children aged 4-6 years were recruited prospectively for clinical validation from June 2020 to December 2020. Conventional single-energy CT (SECT) and SF-DECT were employed to scan the abdominal areas of the phantoms and children. Then, the CT values, image noise, contrast to noise ratios (CNRs), and image subjective scores of SF-DECT and SECT were compared under various doses (1, 2, 3, and 4 mGy).Results For the phantoms under doses of 3 and 4 mGy, SF-DECT decreased the image noise by 18.9% and 23.6%, respectively, and increased the liver and kidney CNRs (CNRliv and CNRkid) by 12.8% and 31.9% at most, respectively, compared to SECT (Z=3.00, 5.17, P < 0.001). For children, SF-DECT decreased image noise (Z=4.64, P < 0.001) and increased CNRliv and CNRkid (Z=3.78, 3.39, P < 0.001). For both the phantoms and the children, the subjective scores of images scanned using the SF-DECT were higher than those scanned using the SECT (Z=1.96-3.80, P < 0.05).Conclusions Compared with SECT, SF-DECT can improve the quality of children's abdominal images. This technique has a certain prospect of optimizing abdominal CT for children. However, it is necessary to conduct in-depth clinical research to verify the result. |