李润根,苏贝贝,马妮,刘莹,白浩.我国CT检查成年人辐射剂量诊断参考水平的建立历程及解读[J].中华放射医学与防护杂志,2020,40(1):71-76
我国CT检查成年人辐射剂量诊断参考水平的建立历程及解读
Evolution and interpretation of diagnostic reference levels for adults undergoing X-ray computed tomography in China
投稿时间:2019-03-23  
DOI:10.3760/cma.j.issn.0254-5098.2020.01.013
中文关键词:  辐射剂量  断层摄影术,X射线计算机  诊断参考水平  辐射剂量优化
英文关键词:Radiation dose  Tomography, X-ray computed  Diagnostic reference levels  Radiation dose optimization
基金项目:
作者单位
李润根 中国兵器工业五二一医院CT室, 西安 710065 
苏贝贝 中国兵器工业五二一医院CT室, 西安 710065 
马妮 1 中国兵器工业五二一医院CT室, 西安 710065 
刘莹 中国兵器工业卫生研究所放射卫生研究室, 西安 710065 
白浩 中国兵器工业五二一医院CT室, 西安 710065 
摘要点击次数: 2544
全文下载次数: 1281
中文摘要:
      经过放射诊断专家、影像技术专家、辐射防护专家和专业技术人员10多年的探索,一项基于大规模国内调查数据而建立的CT检查成年人辐射剂量诊断参考水平(diagnostic reference level,DRL)以国家卫生行业标准(WS/T 637-2018)形式发布。其制定原则和方法符合国际惯例和我国的实际情况,基本上涵盖了我国成年人常见CT检查项目,与国外的DRL比较,整体处于一个合理或较低的剂量水平。给出的50%分位数(可能达到水平)和25%分位数(异常低剂量的提示水平)作为辐射剂量优化指导的额外工具。在日常放射诊断活动中,使辐射剂量与图像质量、临床诊断任务相匹配,降低非正当过高或过低剂量的发生频率。
英文摘要:
      The diagnostic reference level (DRL) for adults radiation dose in CT examination based on a large-scale national survey data is released in the form of national health industry standards (WS/T 637-2018) after more than ten years of exploration by radiologists, imaging technologists, radiation protection specialists and radiographers. Its principles and method are in line with international practices and the actual situation in China, which basically cover frequently-used CT examination items for adults. Compared with DRL in several other countries or organizations, radiation exposure to the patients as a whole is at a reasonably low level. The 50th percentile (achievable dosimetry levels) and 25th percentile (indicative level of unusually low dosimetry) are given as additional tools for radiation dose optimization guidance. In daily activities of radiological diagnosis, the radiation dose should be matched with image quality and clinical diagnostic tasks, and the frequency of unjustifiable high or low radiation dose should be reduced.