贾陈志,魏坤杰,程晓军.河南省调强放疗光子线束吸收剂量和二维剂量分布验证研究[J].中华放射医学与防护杂志,2018,38(2):125-128
河南省调强放疗光子线束吸收剂量和二维剂量分布验证研究
Validation study of photon beam absorbed dose and two-dimensional dose distribution in intensity modulated radiotherapy in Henan province
投稿时间:2017-08-19  
DOI:10.3760/cma.j.issn.0254-5098.2018.02.009
中文关键词:  热释光剂量计  胶片  调强放疗  吸收剂量  二维剂量分布
英文关键词:Thermoluminescent dosimeter  Film  Intensity modulated radiotherapy  Absorbed dose  Two-dimensional dose distribution
基金项目:IAEA基金资助项目(17821/CRP)
作者单位E-mail
贾陈志 450052 郑州, 河南省职业病防治研究院  
魏坤杰 450052 郑州, 河南省职业病防治研究院  
程晓军 450052 郑州, 河南省职业病防治研究院 hnfsws@163.com 
摘要点击次数: 3799
全文下载次数: 1670
中文摘要:
      目的 用热释光剂量计(TLD)和胶片测量调强放疗(IMRT)光子线束吸收剂量和二维剂量分布。方法 采用非概率抽样方法,在河南省选择5家三级甲等医院的8台可开展IMRT的医用加速器,TLD放入国家原子能机构(IAEA)提供的聚苯乙烯固体模体(15 cm×15 cm×15 cm)中,经CT扫描,影像传给放射治疗计划系统(TPS)制定放疗计划,源皮距90 cm,深度10 cm,照射野5 cm×5 cm,6 MV X射线,计算吸收剂量6 Gy和相应的监督单位(MU),实施IMRT计划照射模体,测量TLD吸收剂量,同样方法测量胶片吸收剂量。医院的均质固体模体,尺寸30 cm×30 cm,厚度20 cm,25 cm×25 cm的胶片放在模体中,源皮距95 cm,深度5 cm,实施IMRT计划。结果 调查的8台医用加速器中,有7台加速器的TLD吸收剂量相对偏差符合要求,1台加速器不符合要求;胶片吸收剂量相对偏差全部符合要求;7台加速器的二维剂量分布通过率符合要求,1台加速器不符合要求。结论 TLD和胶片用于核查调强放疗多叶光栅野吸收剂量和二维剂量分布,方法简单,可操作性强,适合在我省医院大范围实施IMRT剂量质量核查。
英文摘要:
      Objective To validate the absorbed dose and two-dimensional dose distribution from photon beam by using Thermoluminescent dosimeter(TLD) and film for intensity modulated radiation therapy (IMRT). Methods A total of 8 medical accelerators were selected among 5 third-grade first-class hospitals using non-probabilistic sampling method in Henan province. TLDs were put into polytetrafluoroethylene solid phantom with size of 5 cm×15 cm×15 cm provided by IAEA. After CT scanning, the radiotherapeutic plans were formulated through image transmission to the treatment planning system (TPS). The IMRT plan was implemented for measurement of TLD-absorbed doses under the conditions of 5 cm×5 cm field, 90 cm SSD, 10 cm depth, 6 MV photon beam and 6 Gy absorbed dose and corresponding measurement unit (MU). The 25 cm×25 cm film-absorbed dose measurement was made in the same manner as TLD under the conditions of 30 cm×30 cm size, 20 cm thickness, 95 cm SSD and 5 cm depth. Results Of eight accelerators, the requirements can be met for 7 accelerators with respect to the relative deviation of TLD absorbed dose except 1. For film, relative deviations were all consistent with the requirements. The passing rate of two-dimensional dose distribution was in line with the requirements for 7 accelerators except 1. Conclusions TLD and film can be used to check the MLC field absorbed dose and two-dimensional dose distribution. This method is simple, easy to operate and suitable for the implementation of IMRT quality control in hospitals in Henan province.
HTML  查看全文  查看/发表评论  下载PDF阅读器
关闭