赵楠,杨瑞杰,王俊杰.热释光探测器的剂量学特性研究[J].中华放射医学与防护杂志,2015,35(9):696-699
热释光探测器的剂量学特性研究
The dosimetric characteristics of thermoluminescent dosimeter
投稿时间:2014-11-17  
DOI:10.3760/cma.j.issn.0254-5098.2015.09.014
中文关键词:  热释光探测器  重复性  剂量响应  剂量率效应  能量响应
英文关键词:Thermoluminescent dosimeter  Repeatability  Dose response  Dose rate effect  Energy response
基金项目:国家自然科学基金(81372420)
作者单位E-mail
赵楠 100191 北京大学第三医院放射治疗科  
杨瑞杰 100191 北京大学第三医院放射治疗科 ruijyang@yahoo.com 
王俊杰 100191 北京大学第三医院放射治疗科  
摘要点击次数: 3113
全文下载次数: 2644
中文摘要:
      目的 研究热释光探测器(TLD)的剂量学特性。方法 选取一致性为±3%的TLD 2000型热释光探测器350片。采用6 MV X射线和125I粒子分别照射,研究热释光探测器对6 MV X射线和125I粒子响应的重复性以及剂量率效应。采用137Cs(662 keV γ射线)、125I粒子、6 MV X射线分别照射热释光探测器,研究热释光探测器的剂量响应。采用125I粒子、137Cs和X射线(48、65、83、118、250 keV和6 MV)对热释光探测器照射相同剂量,研究热释光探测器的能量响应。结果 热释光探测器对于6 MV X射线和125I粒子响应重复性最大偏差分别为2.7%和4.0%,且均无剂量率效应。对于137Cs和125I粒子,热释光探测器的剂量响应均为线性。对于6 MV X射线,剂量线性响应范围为0.74~10 Gy,超过10 Gy热释光剂量响应呈非线性。能量响应按照137Cs进行归一,对于125I粒子,48、65、83、118、250 keV和6 MV能量X射线,能量响应分别为1.70、1.25、1.08、0.99、0.91、0.96和1.22。结论 TLD 2000型热释光探测器的重复性好,对于137Cs、125I粒子和6 MV X射线的剂量响应呈线性,且无剂量率效应,但存在能量响应。
英文摘要:
      Objective To study the dosimetric characteristics of thermoluminescent dosimeter (TLD). Methods A total of 350 pieces of TLD 2000 type TLDs having the same sensitivity within±3.0% were selected. The TLDs were irradiated to study the repeatability and dosage effect of TLDs by using 6 MV X-rays and 125I seeds. In order to study the dose response, the TLDs were irradiated at different dosages by using 137Cs (662 keV γ-rays), 125I seeds and 6 MV X-rays. In order to study the energy response, the TLDs were exposed to the same dose, from 125I seeds, 137Cs and X-rays (48, 65, 83, 118, 250 keV and 6 MV). Results The maximum deviations of the repeatability were 2.7% and 4.0% for 6 MV X-rays and 125I seeds, respectively, and there was no effect of dose rate observed. The dose response of TLDs to 137Cs and 125I seeds were linear. For 6 MV X-rays, the linear response was within the range of 0.74-10 Gy and non-linear range was beyond 10 Gy. The energy response to 125I seeds, 48, 65, 83, 118, 250 keV and 6 MV X-rays, relative to the energy response of 137Cs, were 1.70, 1.25, 1.08, 0.99, 0.91, 0.96 and 1.22, respectively. Conclusions TLD 2000 has a good repeatability and linear dose response for 137Cs, 125I seeds and 6 MV X-rays with no dose rate effects, but the dose response is energy dependent.
HTML  查看全文  查看/发表评论  下载PDF阅读器
关闭