甘家应,胡银祥,罗元强,等.ElektaPrecise直线加速器新型全碳素纤维治疗床床板对放疗剂量的影响[J].中华放射医学与防护杂志,2012,32(4):386-389.GAN Jia-ying,HU Yin-xiang,LUO Yuan-qiang,et al.Effect of the new carbon fiber bed board of Elekta Precise linear accelerator on the radiation dose[J].Chin J Radiol Med Prot,2012,32(4):386-389 |
ElektaPrecise直线加速器新型全碳素纤维治疗床床板对放疗剂量的影响 |
Effect of the new carbon fiber bed board of Elekta Precise linear accelerator on the radiation dose |
投稿时间:2012-02-26 |
DOI:10.3760/cma.j.issn.0254-5098.2012.04.013 |
中文关键词: 治疗床床板 等中心技术 剂量衰减 衰减因子 放疗剂量 |
英文关键词:Treatment tabletop Source-axis distance technology Dose attenuation Attenuation factor Radiotherapy dose |
基金项目:贵州省科学技术基金[黔科合(200)号2059] |
|
摘要点击次数: 4050 |
全文下载次数: 2617 |
中文摘要: |
目的 探讨Elekta Precise直线加速器新型全碳素纤维治疗床床板对放疗剂量的影响。方法 应用等中心技术(SAD)测量,设2组射野,均为等中心对穿野(一组穿过床板,另一组不穿过床板),用PTW剂量仪0.6 cm3 Farmer电离室比对测量,计算出治疗床主床板、体部延长板、头颈肩延长板不同位置及其衔接处对放疗剂量的衰减。结果 6 MV能量时,治疗床主床板对剂量的衰减为1.4%~7.2%,主床板衔接头1、4、7和8 cm处对剂量的衰减分别为2.8%~38.7%、1.4%~30.1%、1.5%~20.8%和1.4%~11.2%;体部延长板对剂量的衰减为0.5%~5.0%,体部延长板衔接头1 cm位置对剂量的衰减为4.7%~15.4%;头颈肩延长板颈部位置对剂量的衰减在0.5%~3.3%,头颈肩延长板肩部位置对剂量的衰减在5.3%~16.7%;延长板和主床板衔接处对剂量的衰减在6.8%~30.4%。结论 新型全碳素纤维治疗床床板不同部位对剂量衰减不同,床板衔接部位标记区域对剂量衰减比较大。 |
英文摘要: |
Objective To investigate the dosimetric influence of pure carbon fiber treatment tabletop of Elekta Precise new linear accelerator in radiotherapy.Methods Surface-axis distance (SAD) technology was employed for the measurement. Two groups of fields were set and both of them were SAD opposed portals (one of them went through the tabletop, while the other did not). A PTW electrometer and a 0.6 cm3 Farmer ionization chamber were utilized for comparison measurement. Then dose attenuation of the main table board, extended body board, the extended board for head, neck and shoulders, and the joints of these boards were calculated. Results Under the energy of 6 MV, the dose attenuations of the following locations were: 1.4%-7.2% at the main treatment table board; 2.8%-38.7%, 1.4%-30.1%, 1.5%-20.8% and 1.4%-11.2%, respectively at distances of 1, 4, 7 and 8 cm from the joint of the main table board;0.5%-5.0% at the extended body board; 4.7%-15.4% at distance of 1cm from the joint of the extended body board; 0.5%-3.3% at the neck position of the extended board for head, neck and shoulders; 5.3%-16.7% at the shoulder positions; and 6.8%-30.4% at the joint between the extended boards and the main table board. Conclusions The dose attenuations of the new linear accelerator pure carbon fiber treatment tabletop vary at different locations. Considerable higher attenuations are observed at the table board joints than other locations. |
HTML 查看全文 查看/发表评论 下载PDF阅读器 |
关闭 |
|
|
|