武大鹏,万玲,娄云,等.脑血管造影术中X射线机房内辐射剂量分布研究[J].中华放射医学与防护杂志,2011,31(4):387-390.WU Da-peng,WAN Ling,LOU Yun,et al.Distribution of radiation dose in X-ray room in cerebral angiography operations[J].Chin J Radiol Med Prot,2011,31(4):387-390
脑血管造影术中X射线机房内辐射剂量分布研究
Distribution of radiation dose in X-ray room in cerebral angiography operations
投稿时间:2011-02-17  
DOI:10.3760/cma.j.issn.0254-5098.2011.04.003
中文关键词:  介入放射学  剂量率  辐射场
英文关键词:Interventional radiology  Dose rate  Radiation field
基金项目:国家科技支撑计划(2006BAI06B03)
作者单位E-mail
武大鹏 100013 北京市疾病预防控制中心  
万玲 100013 北京市疾病预防控制中心  
娄云 100013 北京市疾病预防控制中心 louyun118@263.net 
冯泽臣 100013 北京市疾病预防控制中心  
王新明 100013 北京市疾病预防控制中心  
侯磊 吉林大学公共卫生学院  
王宏芳 吉林大学公共卫生学院  
王树华 北京医院  
彭建亮 100013 北京市疾病预防控制中心  
翟曙光 100013 北京市疾病预防控制中心  
摘要点击次数: 5129
全文下载次数: 3288
中文摘要:
      目的 研究脑血管造影术中不同照射方向下X射线机房内辐射剂量场的分布情况。方法 选用脑血管造影术程序,选取后前位、前后位、右前斜30?、左前斜30?、右侧位和左侧位6种照射方式,使用451B型空气电离室巡测仪对以散射模体为中心、半径为3 m范围内的辐射场的剂量率分布进行测量。结果 后前位及右侧位照射时主要操作人员所受到的辐射剂量最低,均为0.4 mGy/h;左侧位照射时主要操作人员受到的辐射剂量最高,为1.54 mGy/h;机架侧区域的辐射剂量水平最低,均小于10 μGy/h。结论 在脑血管造影术中,医护操作人员应尽可能选择剂量较低的后前位及右侧位照射方式投照,并合理选择工作位置,可明显降低受照剂量。
英文摘要:
      Objective To determine the distribution of irradiation dose in X-ray room with different radiation directions in cerebral angiography operations. Methods The Model 451B air ionization chamber survey meter was used to measure the irradiation dose distribution in the X-ray room when cerebral angiographies were performed at high kV and high dose. The doses were measured within 3 m when the phantom was irradiated in posterior-anterior (PA) position, anterior-posterior(AP) position, right anterior oblique (RAO) 30?, left anterior oblique (LAO) 30?, right-lateral (RL) position, and left-lateral(LL) position, respectively. Results Both the relative data and the distribution curves of the six radiation directions were obtained by using the same interventional operation model. The irradiation dose to the primary operator was minimal in P-A and RL positions,at 0.4 mGy/h, and was maximal in LL position, at up to 1.54 mGy/h. Furthermore, the lowest irradiation dose level was found to be lower than 10 μGy/h at the side of the framework field. Conclusions Low-irradiation position and rational work place were proposed to reduce the operators’ absorption dose during interventional radiology procedures.