钱玉娥,胡红杰,王丹,王高峰,江峰,章士正.二代双源CT对冠状动脉支架的显示及辐射剂量的研究[J].中华放射医学与防护杂志,2011,31(3):366-369
二代双源CT对冠状动脉支架的显示及辐射剂量的研究
Assessment of coronary artery stent and radiation dose using dual source CT
投稿时间:2011-01-16  
DOI:10.3760/cma.j.issn.0254-5098.2011.03.033
中文关键词:  双源CT  冠状动脉  支架  辐射剂量
英文关键词:Dual source CT  Coronary artery stent  Radiation dose
基金项目:
作者单位
钱玉娥 310016 杭州,浙江大学医学院附属邵逸夫医院放射科 
胡红杰 310016 杭州,浙江大学医学院附属邵逸夫医院放射科 
王丹 310016 杭州,浙江大学医学院附属邵逸夫医院放射科 
王高峰 310016 杭州,浙江大学医学院附属邵逸夫医院放射科 
江峰 310016 杭州,浙江大学医学院附属邵逸夫医院放射科 
章士正 310016 杭州,浙江大学医学院附属邵逸夫医院放射科 
摘要点击次数: 2804
全文下载次数: 2051
中文摘要:
      目的 对比分析二代双源CT前瞻性心电触发高螺距螺旋扫描模式(Flash模式)和回顾性心电门控螺旋扫描模式(常规模式)下冠状动脉支架的显示质量和辐射剂量。方法 常规模式和Flash模式下共计120例患者的155个冠状动脉植入支架纳入本研究,平均年龄(64.9±10.6)岁,所有患者的心率均控制在≤65次/min,且心律规则。所有患者分成两组,分别接受Flash模式和常规模式的冠状动脉成像。冠状动脉原始图像均进行Kernel B26和Kernel B46的数据重建,并由两位医师对支架的显示质量进行独立的4级评分。两种模式下的冠状动脉显示质量利用卡方检验进行统计分析,而不同冠状动脉扫描模式下的容积CT剂量指数值(CTDIvol)和剂量长度乘积值(DLP),利用t检验进行辐射剂量的对比研究。结果 2位医师对所有冠状动脉支架显示质量的评分具有较好的一致性( Kappa =0.764, P <0.001),Flash模式和常规模式的平均评分分别为:1.61±0.77 和1.65±0.82。两种冠状动脉扫描模式的图像质量差异无统计学意义( χ2 =0.865, P =0.834)。Flash模式的CTDIvol(3.24±1.21)明显低于常规模式(31.26±10.79),差异有统计学意义(t=19.83, P <0.001);Flash模式的DLP(54.61±19.88)同样低于常规模式(468.30±174.88),差异有统计学意义(t=18.06, P <0.001)。结论 控制患者心率≤65次/min且心律规则,Flash模式在降低辐射剂量的前提下可获得与常规模式相同质量的冠状动脉支架图像。
英文摘要:
      Objective To compare a prospective ECG-gated high-pitch spiral technique (Flash) and conventional retrospective ECG-gated spiral technique for the image quality of coronary artery stent and radiation dose with a dual source CT. Methods One hundred and fifty five coronary stents in one hundred and twenty patients (mean age 64.9±10.6 years, heart rates≤65 bpm) were examined using a dual source CT. All patients were divided in two groups, receiving either Flash or conventional coronary artery CT angiography separately. After images of coronary artery were reconstructed using both the smooth (B26) and sharp (B46) kernel, the coronary stent image quality and stent lumen were scored by two observers individually using four point scale (1=excellent, 4=unvaluable). The effective radiation dose of volume CT dose index (CTDIvol, mGy) and dose length product (DLP, mGy×cm) were also calculated for each patient. χ2 -test analysis of image quality andt-test analysis of radiation dose were used respectively for statistical difference between two groups. Results Interobserver agreement for stent image quality was good ( Kappa =0.764, P <0.001). The mean scores were 1.61±0.77 and 1.65±0.82 in Flash group and conventional group respectively. There was no significant difference in image quality between the two groups ( χ2 =0.865, P =0.834). The effective radiation dose in Flash group was significantly lower than that in conventional group. The mean values of CTDIvol were 3.24±1.21 in Flash group and 31.26±10.79 in conventional group (t=19.83, P <0.001), and the mean values of DLP in Flash group and conventional group were 54.61±19.88 and 468.30±174.88, respectively (t=18.06, P <0.001). Conclusions Compared with the conventiaonal coronary artery CT angiography, the Flash coronary artery CT angiography technique has a similar coronary stent image quality, but at a lower radiation dose in patients with heart rates lower than 65 beats per minute.
HTML  查看全文  查看/发表评论  下载PDF阅读器
关闭