马恩森,王武,马国林,杨志刚,栾辉.低剂量64层螺旋CT测定冠状动脉钙化积分的准确性评价[J].中华放射医学与防护杂志,2010,30(5):605-609
低剂量64层螺旋CT测定冠状动脉钙化积分的准确性评价
Accuracy of coronary calcium measurements with low dose 64-slice multi-detector row CT
投稿时间:2010-06-02  
DOI:
中文关键词:  冠状动脉粥样硬化性心脏病  体层摄影术,X线计算机  钙化沉着症
英文关键词:Coronary artery disease  Tomography, X-ray computed  Atherosclerosis  Calcium
基金项目:
作者单位E-mail
马恩森 100029 北京, 中日友好医院放射科 Mgl4022005@yahoo.com.cn 
王武 100029 北京, 中日友好医院放射科  
马国林 100029 北京, 中日友好医院放射科  
杨志刚 100029 北京, 中日友好医院放射科  
栾辉 四川大学华西医院放射科  
摘要点击次数: 2856
全文下载次数: 2650
中文摘要:
      目的 探讨低剂量64层螺旋CT测定冠状动脉钙化积分的准确性。方法 2006年8月至2009年9月间对43名冠状动脉钙化的患者连续进行2次64层螺旋CT扫描,管电流时间积分别为常规剂量(100 mAs)和低剂量(55 mAs),其余参数不变,由2名放射科副主任医师测定钙化积分并测量升主动脉CT值的均数及标准差。结果 低剂量与常规剂量钙化总积分和独立血管钙化积分取平方根转换后均直线相关(r =0.998、0.997)。低剂量管电流时间积所测升主动脉根部的CT值的均数与2倍标准差之和小于130 HU,可满足所要求的信噪比,与常规剂量扫描辐射剂量(1.32±0.08)mSv相比较,低剂量扫描有效剂量降低0.6 mSv(P <0.05)。结论 低剂量前瞻性心电门控64层MDCT钙化积分扫描图像能满足测量需要,测量结果具有较高的准确性。
英文摘要:
      Objective To compare the low- and standard-dose protocols in the evaluation of coronary artery calcification (CAC) scores and in image quality, and to assess the accuracy of low-dose prospective ECG-gated helical CT in CAC measurements. Methods Forty-three patients with CAC were scanned twice using a tube current setting of 100 and 55 mA retrospectively. CAC scores (Agatston) of different branches of coronary artery were recorded. The mean and SD of the CT attenuation values in regions of interest in the aorta were measured, and the value (mean+2×SD) were obtained. Results Calcification plaques were found in 105 of the 172 vessels in the 43 subjects. The radiation dose of the 50 mAs group was (0.72±0.04) mSv, significantly lower by 0.6 mSv than that of the 100 mAs group [(1.32±0.08)mSv, P <0.05]. The total calcification score of the 55 mAs group was significantly positively correlated with the routine dose group after performing the square root transformation (r =0.998). A signficant correlation existed in vessel-based calcification between these 2 groups (r = 0.997). Conclusions Low-dose prospective ECG-gated 64-slice CT yields high accuracy on CAC scoring while maintaining the image quality needed to measure CAC.
HTML  查看全文  查看/发表评论  下载PDF阅读器
关闭