李建明,徐文彪,罗源利,刘鸿圣,周宁.64层螺旋CT自动管电流调节技术对婴幼儿头颅扫描的应用[J].中华放射医学与防护杂志,2010,30(2):222-224
64层螺旋CT自动管电流调节技术对婴幼儿头颅扫描的应用
Application of automatic tube current modulation with 64-detector CT in infant brain
投稿时间:2009-06-08  
DOI:
中文关键词:  自动管电流调节技术  婴幼儿  头颅  低剂量
英文关键词:Automatic tube current modulation  Infant  Brain  Low dose
基金项目:
作者单位
李建明 510120 广州市儿童医院放射科 
徐文彪 510120 广州市儿童医院放射科 
罗源利 510120 广州市儿童医院放射科 
刘鸿圣 510120 广州市儿童医院放射科 
周宁 510120 广州市儿童医院放射科 
摘要点击次数: 2787
全文下载次数: 2048
中文摘要:
      目的 评估64层螺旋CT自动管电流调节技术对婴幼儿头颅的图像质量和辐射剂量的影响。 方法 按2.5、2.75、3.0和4.0的SD(标准差)值将200例婴幼儿随机分成4组,用自动管电流调节技术行头颅CT扫描,每组各50例,回顾性统计分析各组的图像质量和辐射剂量。结果 当SD为2.5、2.75、3.0和4.0时,CTDIvol最高值分别为60、49.8、42.9和25.8 mGy,有效剂量分别为(3.27±1.01)、(2.78±0.85)、(2.40±0.74)和(1.49±0.45)mSv·mGy-1·cm-1,各组比较差异均有统计学意义(F=48.99, P<0.05);图像质量优秀率分别为100%、96%、70%和20%,其中SD 2.5组新生儿12例,SD 2.75组新生儿(2例)的图像质量均为良,SD 3.0组图像质量良以下均为<1周岁的病例,SD 4.0组图像质量优均为>2岁6个月和头围较大的病例。结论 ATCM的个性化应用可实现婴幼儿头颅CT的图像质量、辐射剂量和诊断图像的最佳平衡,建议对新生儿、1个月~1周岁、1~3周岁和头围较大或复查病例分别采用2.5、2.75、3.0和4.0 的SD值。
英文摘要:
      Objective To evaluate the effect of automatic tube current modulation on imaging quality and lesion revealed in infant brain.Methods The infant patients of 200 cases were divided into four groups according to the values of SD(2.5, 2.75, 3.0, 4.0), 50 cases in each group. They were performed with the automatic tube current modulation scanning in brain. The imaging quality and radiation dose were analyzed.Results When the values of SD were 2.5, 2.75, 3.0 and 4.0, the high values of CTDIvol were 60.0, 49.8, 42.9 and 25.8 mGy, respectively. The effective doses were (3.27±1.01), (2.78±0.85), (2.40±0.74) and (1.49±0.45) mSv·mGy-1·cm-1, respectively. There was significant difference among those groups (F =48.99, P<0.05). The excellent imaging qualities were 100%, 96%, 70% and 20%. 12 cases were neonate in SD 2.5 group. In SD 2.75 group, the imaging quality of neonate(2 cases)were all fine. In SD 3.0 group, the imaging quality was not ideal for the ages less than 1 year. In SD 4.0 group, when the infants aged over 2 years and 6 months, or with the larger head circumference, their imaging quality were all excellent.Conclusions The individual application of ATCM could obtain the best balance among imaging quality, radiation dose and diagnostic quality. The SD 2.5, 2.75, 3.0 and 4.0 might be suitable to infants of newborn, 1 month to 1 year old, 1 to 3 years, the larger head circumference and part of the reviewed cases.
HTML  查看全文  查看/发表评论  下载PDF阅读器
关闭