贾明轩,邹华伟,纪天龙,张旭,韩成波.兆伏级锥形束CT图像引导放疗所致鼻咽癌患者剂量的探讨[J].中华放射医学与防护杂志,2010,30(2):199-201
兆伏级锥形束CT图像引导放疗所致鼻咽癌患者剂量的探讨
Evaluation of the patient doses from megavoltage cone-beam CT imaging in the image-guided radiation therapy for nasopharyngeal carcinoma.
投稿时间:2009-04-17  
DOI:
中文关键词:  鼻咽癌  兆伏级锥形束CT  患者剂量
英文关键词:Nasopharyngeal carcinoma  Megavoltage cone-beam CT  Patient doses
基金项目:
作者单位
贾明轩 110022 沈阳,中国医科大学附属盛京医院肿瘤治疗中心 
邹华伟 110022 沈阳,中国医科大学附属盛京医院肿瘤治疗中心 
纪天龙 110022 沈阳,中国医科大学附属盛京医院肿瘤治疗中心 
张旭 110022 沈阳,中国医科大学附属盛京医院肿瘤治疗中心 
韩成波 110022 沈阳,中国医科大学附属盛京医院肿瘤治疗中心 
摘要点击次数: 2827
全文下载次数: 2077
中文摘要:
      目的 评价和估算兆伏级锥形束CT(MV CBCT)成像系统在图像引导放疗中所致鼻咽癌患者的辐射剂量。方法 选择MV CBCT系统头颈部8 MU扫描预案,利用0.65 cm3指型电离室和CT头部剂量体模测量出体模不同位置的吸收剂量。并利用XiO治疗计划系统模拟MV CBCT扫描过程,计算体模电离室测量点的吸收剂量和鼻咽癌患者肿瘤靶区及危及器官的吸收剂量。结果 体模不同位置吸收剂量的测量值和计算值具有很好的一致性,相对误差均小于3.5%。MV CBCT图像引导放疗所致鼻咽癌患者肿瘤靶区平均剂量为6.43 cGy,脑干、脊髓和视交叉的平均剂量分别为6.36 、6.83和6.90 cGy,左、右视神经平均剂量分别为7.70和7.53 cGy,左、右腮腺平均剂量分别为6.86和6.43 cGy。结论 使用治疗计划系统模拟MV CBCT图像采集过程估算剂量准确、可靠。在设计患者治疗计划时,要充分考虑MV CBCT图像采集过程所致患者剂量。
英文摘要:
      Objective To evaluate and estimate the patient doses from megavoltage cone-beam CT imaging system in the image-guided radiation therapy for nasopharyngeal carcinoma.Methods 8 MU protocol of the MV CBCT system was selected for the head-and-neck region. The absorbed doses at the different positions in the phantom were measured using a 0.65 cm3 ion chamber and the cylindrical acrylic phantom. The absorbed doses at the measurement positions of the phantom were calculated and the patient doses to the tumor and critical organs were derived from dose-volume histogram by the TPS mimicking the MV CBCT scanning with 8 MU protocol.Results The error between the measured dose and the calculated dose was less than 3.5%. The average doses to the tumor target, brain stem, spinal cord and chiasm were 6.43, 6.36, 6.83 and 6.90 cGy, respectively, while those to left and right of both optic nerve and parotid were 7.70 and 7.53 cGy, 7.70 and 7.53 cGy, respectively.Conclusion The patient doses estimated using the TPS mimicking the MV CBCT image acquiring procedure are accurate and reliable. The patient doses from the MV CBCT imaging must be considered when treatment plan of the patient is designed.
HTML  查看全文  查看/发表评论  下载PDF阅读器
关闭